Ever since the technique to blot protein
samples on nitrocellulose membrane was developed, Western blot has been a go-to
technique for every molecular biologist. The method is simple enough and the
results seem straightforward, but correct interpretation of your results is a
crucial and often overlooked step.
In Figure 1, I present an example of a
straightforward Western blot result. For this experiment, I differentiated the
myeloid cell line PLB into functioning granulocytes by stimulating them with dimethylformamide
(DMF) for 6 days . Afterwards, I
took a sample of my stimulated and control cells as well as a sample of primary
human granulocytes (PMN), dissolved these in sample buffer, ran the samples
next to each other on an SDS-PAGE gels, blotted the gel to a membrane and
probed that membrane with antibodies against the NADPH-oxidase components p67
and p47. Finally, I probed the blot with fluorescently labelled antibodies and
scanned it on a Li-Cor Odyssey infrared scanner to detect the signal. As you
can see, the neutrophils (PMN) express high levels of both p67 (in red) and p47
(in green) whereas the control PLB cells express nothing and the treated PLB
cells express levels comparable to the neutrophils, demonstrating that the
experiment was successful. I also probed the blot for β-actin to demonstrate
equal loading of the samples. This is a perfectly straightforward result: the
protein is either there or not there at all. No discussion about it! However,
most blots won’t be so straightforward.
In most cases, your protein of interest
won’t be either there or not there. The level of expression will be reduced or
increased by a certain margin. Western blots can be very deceptive in
demonstrating such subtle effects, as I will outline below.
Controls
First and foremost in scientific practice
come the controls. Your control and experiment samples should have received the
exact same treatment to be valid and run side by side on a gel for proper
comparison. I know it happens far too often that you go through the whole
procedure only to discover a nasty spot in either your sample or control lane.
Then you go and repeat the whole procedure, only to have the other lane smudged
this time. If only you cropped both pictures and spliced them next to each
other, you’d have a perfect result… Unfortunately, this is not allowed. Such
practice reeks of scientific fraud and makes it impossible to judge whether or
not your result was real, even if you mean well. Nope, my friend, either it’s
back to the lab, or show the smudges! There’s nothing wrong with showing
smudges, one should not be afraid of that. Smudges are much preferred over
fraud!
Detection
limits
The second issue to consider before stating
whether or not a certain protein has disappeared or not (for example after
siRNA treatment) is the detection limit of your protein of interest. The
problem with all commercial antibodies is that nobody really knows just how
strongly they interact with their target protein. Anybody who’s tried several
different antibodies against the same target knows that some will give you a
very strong signal and low background, while others will give you a weak signal
and high background. Anything in between, in any possible combination, can also
be encountered. It’s completely random! This indicates that the strength of the
signal on your blot is not so much an indication of relative protein abundance
as of antibody quality. Though, of course, protein abundance also helps.
Detection limit becomes especially relevant
when your signal is very weak. This means you’re drawing conclusions close to
the detection limit of your assay. Thus, when comparing two lanes on a blot, it
may seem that in one lane the protein is there and in the other it’s absent,
while in reality the difference is no more than 10% or so. What makes the blot
deceptive is the detection limit of your protein. It could just happen that you
loaded 10% less of your experimental sample and that this is enough to make
your protein of interest undetectable, even though it’s still there. If you now
compare the control and your experimental lane, it will look like your protein
of interest has disappeared completely after treatment and since we usually use
highly abundant proteins, such as actin, as our loading controls, this 10%
difference in loading will go unnoticed. I see these kind of results quite
often with siRNA controls; very weak signals that appear to show a black and
white difference. Results collected on old fashioned X-ray film are
particularly sensitive to this form of deception, since a pixel on the film is
either black or not. The grey values on a film are entirely derived from pixel
density. In contrast, results collected with a fluorescence imager are much
less sensitive to deception, since every pixel can have a wide range of values.
On the Li-Cor Odyssey, for example, a pixel can have a value between 1 and
40,000, providing a huge dynamic range.
Post-processing
errors
Every digital image can be manipulated to
show only what one desires others to see. Of course, manipulating only part of
the image (say the control lane) and not the rest is fraud, but there’s a huge
grey area of manipulation that is allowed, but not quite correct. I provide an
example in Figure 2. For this experiment, I tried to knock down a gene with
siRNA. The blot shows my control sample (Ctrl) and two different siRNA’s (1 and
2). Panel A is what the blot actually looks like after I acquired the data with
the Odyssey infrared imager, panel B is manipulated to show what I want to see
(protein levels decreased after siRNA treatment) and panel C is my loading
control (HSP90). As you can see, the effect of siRNA treatment appears to be
much greater in panel B than in panel A, even though both show the exact same
blot. What I did to generate panel B was to adjust the image display curves,
rather than merely the contrast. You should never, ever do that! Ever. Because
by adjusting the curves you’re discarding data you don’t like. You’re telling
the program that you don’t care about values above or below a certain threshold,
thus you get rid of pixels with very low or very high values. Of course, this
eliminates your background, but you also lose information that might be
valuable, such as the weak bands visible in panel A below and above the main
bands. In addition, you enhance and multiply a small difference to make it seem
much greater. This might indeed help you get your work published in Nature, wherein most blots are
suspiciously squeaky clean, but it really isn't the way to go.
Figure 2. Control sample (Ctrl) and samples treated with siRNA 1 or 2. A) unprocessed image, B) processed image, C) loading control (HSP90). |
Loading
controls
Another obvious problem is with the loading
control. As I mentioned before, we tend to choose a highly abundant protein as
our loading control, such as Actin. However, if your control is much more
abundant than your protein of interest, the result may be highly deceptive. I
provide an example in Figure 3. For this experiment, I simply took a cell
lysate and diluted this with sample buffer in 10%-steps (thus, lane 1 is 100%,
lane 2 90%, 80% etc.). Then I probed the blot with an anti-actin antibody,
followed by a fluorescently-labelled antibody and I scanned the blot with the
Odyssey Infrared scanner. The nice thing about this technology is that I can
actually quantify my signal and draw a graph, as shown in the figure. Now, you’ll
notice that the difference between the first 5 lanes is very hard to spot with
the naked eye, even though there’s an almost 2-fold difference in loading (100%
vs 60%)! In addition, the computer can barely tell the difference between the
first 3 lanes, even though I loaded only 80% of the sample in lane 3. Thus,
also the intensity of your loading control can be very deceptive and you should
be aware that the same phenomenon occurs for every protein you blot for. If you’re
trying to draw conclusions outside the linear range of detection, whether by
eye or computer, you’re going to be deceived.
Figure 3. Decreasing amounts of sample were loaded on gel with steps of 10%. The blot was probed for human Actin and data acquired on the Li Cor Odyssey. |
Finally, the molecular weight of your
controls and experimental samples matter. Not every protein in your sample is
going to transfer equally well and molecular weight is an important determinant
for transfer efficiency. Large proteins tend to transfer slower whereas small
proteins can actually be transferred straight through your blot if your
transfer time is too long. Thus, ideally, your loading control and protein of
interest should be of both similar size and similar abundance.
Solutions
It’s up to you to provide the right
controls for your experiments and to make sure your samples and controls are on
the same blot. You can not, and should not, compare samples from different
blots as they may have been differently exposed and processed.
You should know the detection limit of your
protein of interest and the affinity of your antibody. When you first start
using a new antibody, run a control blot with a dilution range of your sample
to make sure you’re measuring in the linear range of your protein.
Image manipulation is allowed, of course,
you’re already doing it when you’re exposing your film for different lengths of
time. However, only manipulate the whole blot and never discard data. Better to
show some noise and background. Find some useful hints and tips on this website. Nowadays, I much prefer using fluorescence
scanners, such as the Odyssey, the acquisition Western blot data because the
scanned imaged can be quantified and I can easily make different exposures of
my blot, simply by setting the exposure level. When using the Odyssey, always
set the acquisition for the maximal possible exposure, without over-exposing
the image (shown as white pixels in the colour view or blue pixels in the grey
scale view). That way, you collect as much data as possible and you can always
post-process your image to get a prettier picture. More important than a pretty
picture, however, is the ability to graph and calculate your data.
I hope these examples and guidelines give
you some idea of how to interpret Western blot results, be it your own data or
published work.
No comments:
Post a Comment